Reproducibility of Natural Shear Wave Elastography Measurements

Ultrasound Med Biol. 2019 Dec;45(12):3172-3185. doi: 10.1016/j.ultrasmedbio.2019.09.002. Epub 2019 Sep 26.

Abstract

For the quantification of myocardial function, myocardial stiffness can potentially be measured non-invasively using shear wave elastography. Clinical diagnosis requires high precision. In 10 healthy volunteers, we studied the reproducibility of the measurement of propagation speeds of shear waves induced by aortic and mitral valve closure (AVC, MVC). Inter-scan was slightly higher but in similar ranges as intra-scan variability (AVC: 0.67 m/s (interquartile range [IQR]: 0.40-0.86 m/s) versus 0.38 m/s (IQR: 0.26-0.68 m/s), MVC: 0.61 m/s (IQR: 0.26-0.94 m/s) versus 0.26 m/s (IQR: 0.15-0.46 m/s)). For AVC, the propagation speeds obtained on different day were not statistically different (p = 0.13). We observed different propagation speeds between 2 systems (AVC: 3.23-4.25 m/s [Zonare ZS3] versus 1.82-4.76 m/s [Philips iE33]), p = 0.04). No statistical difference was observed between observers (AVC: p = 0.35). Our results suggest that measurement inaccuracies dominate the variabilities measured among healthy volunteers. Therefore, measurement precision can be improved by averaging over multiple heartbeats.

Keywords: Elastography; High frame rate; Natural shear wave elastography; Shear waves; Tissue elasticity imaging; Valve closure; stiffness.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aortic Valve / anatomy & histology*
  • Aortic Valve / physiology
  • Elasticity Imaging Techniques / methods*
  • Female
  • Healthy Volunteers
  • Humans
  • Male
  • Middle Aged
  • Mitral Valve / anatomy & histology*
  • Mitral Valve / physiology
  • Reference Values
  • Reproducibility of Results
  • Young Adult