Radiation dosimetry considerations for skeletal survey imaging of multiple myeloma

Phys Med. 2019 Aug:64:109-113. doi: 10.1016/j.ejmp.2019.06.012. Epub 2019 Jul 10.

Abstract

Purpose: This paper aims to review the dosimetry and utility of currently implemented imaging modalities for assessment of multiple myeloma and consider the role of tin filtration computed tomography (CT) as a potential replacement to current standard practice.

Methods: Radiation output of tin CT was measured experimentally and used for software-based dose calculation. Resultant effective dose was then compared to calculated planar radiography doses and published doses of other imaging modalities.

Results: Based on example patient parameters used for modalities and 14 projection planar radiography site protocols, doses are comparable between planar radiography and tin filtration CT (approximately 0.9 and 1.0 mSv respectively). Both studies carried a reduced radiation burden compared to Expected Pathologically Increased Contrast-CT (EPIC-CT), FDG-PET and MIBI SPECT/CT (5.7, 11.1-20.0 and 13.0 mSv respectively).

Conclusion: Tin filtered CT provided visualisation of multiple myeloma at doses comparable to planar radiography and where available may be a suitable alternative, following due consideration of patient specific justification and optimisation in line with best practice.

Keywords: CT; Dosimetry; Filtration; Myeloma; Tin.

MeSH terms

  • Bone and Bones / diagnostic imaging*
  • Humans
  • Image Processing, Computer-Assisted
  • Multiple Myeloma / diagnostic imaging*
  • Positron-Emission Tomography
  • Radiation Dosage
  • Radiometry
  • Single Photon Emission Computed Tomography Computed Tomography*
  • Software