Influence of Variation in Sagittal Placement of the Femoral Component after Cruciate-Retaining Total Knee Arthroplasty

J Knee Surg. 2021 Mar;34(4):444-451. doi: 10.1055/s-0039-1696958. Epub 2019 Sep 9.

Abstract

Prosthetic alignment is an important factor for long-term survival in cruciate-retaining (CR) total knee arthroplasty (TKA). The purpose of this study is to investigate the influence of sagittal placement of the femoral component on tibiofemoral (TF) kinematics and kinetics in CR-TKA. Five sagittal placements of femoral component models with -3, 0, 3, 5, and 7 degrees of flexion are developed. The TF joint kinematics, quadriceps force, patellofemoral contact force, and posterior cruciate ligament force are evaluated using the models under deep knee-bend loading. The kinematics of posterior TF translation is found to occur with the increase in femoral-component flexion. The quadriceps force and patellofemoral contact force decrease with the femoral-component flexion increase. In addition, extension of the femoral component increases with the increase in posterior cruciate ligament force. The flexed femoral component in CR-TKA provides a positive biomechanical effect compared with a neutral position. Slight flexion could be an effective alternative technique to enable positive biomechanical effects with TKA prostheses.

MeSH terms

  • Arthroplasty, Replacement, Knee / methods*
  • Biomechanical Phenomena / physiology
  • Computer Simulation*
  • Finite Element Analysis*
  • Humans
  • Knee Joint / physiology*
  • Knee Joint / surgery