[Design and Verification of a COPD Screening Equipment Based on Dual-differential Pressure Throttling Technique]

Zhongguo Yi Liao Qi Xie Za Zhi. 2019 Jul 30;43(4):263-265. doi: 10.3969/j.issn.1671-7104.2019.04.008.
[Article in Chinese]

Abstract

This study was to design a chronic obstructive pulmonary disease (COPD) screening equipment, based on the dual-differential pressure throttling technique. The technique combined a wide range, but low-resolution ratio sensor and a narrow range, but high-resolution ratio one. It can accurately detect the indexes of forced vital capacity (FVC), forced expiratory volume in one second (FEV1), one second rate(FEV1/FVC (%)), and achieve them in a low-cost way. The new designed machine will be compared with a British machine, named ML-3500. The correlations of FVC and FEV1 between new machine and ML-3500 were 0.998 and 0.999, respectively. The P values of paired t test of these two indexes were over 0.05. Bland-Altman analysis of FVC, FEV1 and FEV1/FVC (%) showed that more than 90% of the scatter points of the three parameters fell within the consistency interval. This machine can be used to accurately screen COPD and its low-cost would be advantage to promote in large population.

Keywords: COPD; FEV1; FVC; dual-differential pressure throttling technique.

MeSH terms

  • Forced Expiratory Volume
  • Humans
  • Pulmonary Disease, Chronic Obstructive* / diagnosis
  • Reproducibility of Results
  • Respiratory Function Tests* / economics
  • Respiratory Function Tests* / instrumentation
  • Tidal Volume
  • Vital Capacity