Biomechanical properties of off-axis screw in Pauwels III femoral neck fracture fixation: Bicortical screw construct is superior to unicortical screw construct

Injury. 2019 Nov;50(11):1889-1894. doi: 10.1016/j.injury.2019.07.020. Epub 2019 Aug 10.

Abstract

Objectives: The purpose of this study is to determine the biomechanical properties of the bicortical off-axis screw fixation for stabilizing of Pauwels III femoral neck fractures compared with other fixation methods.

Methods: Eighteen synthetic femurs (Sawbones Pacific Research Laboratories, Vashon, WA) were divided into three groups. The osteotomy was made vertically to mimic the Pauwels type III femoral neck fracture. Group A (n = 6) was fixed with traditional inverted triangle cannulated screws. Group B (n = 6) was fixed with a unicortical off-axis screw and two parallel cannulated screws. Group C (n = 6) was fixed with a bicortical off-axis screw and two parallel cannulated screws. Each group was tested with a nondestructive axial compression test at a 7° of valgus followed with 1000 cycles of cyclic loading test from 100 N to 1000 N. Finally, a destructive axial compression test was applied until catastrophic failure.

Results: The average axial stiffness from group A to group C was 856.5, 934, and 1340 N/mm, respectively. The average ultimate failure load from group A to group C was 2612.7, 2508.8, and 3706 N, respectively. Group C exhibited significantly greater axial stiffness and a higher ultimate failure load than the other two groups (P < 0.05). Regarding the interfragmental displacement, the values from group A to group C were 0.41, 0.83, 0.36, respectively, and group B exhibited significantly larger fracture gap formation after the cyclic loading test.

Conclusions: The results of this biomechanical study show statistically significant increases in axial stiffness and ultimate failure load for the off-axis screw placed in bicortical fashion. Once the off-axis screw was positioned unicortically, the largest fracture diastasis was observed as compared to the other two methods.

Keywords: Biomechanics; Fracture fixation; Off-axis screw; Vertical femoral neck fracture.

MeSH terms

  • Artificial Organs
  • Biomechanical Phenomena
  • Bone Screws*
  • Bone and Bones
  • Femoral Neck Fractures / surgery*
  • Fracture Fixation, Internal* / instrumentation
  • Fracture Fixation, Internal* / methods
  • Humans
  • Materials Testing
  • Models, Biological
  • Simulation Training
  • Stress, Mechanical
  • Torque