Optimizing the Analytical Performance of Substrate-Integrated Hollow Waveguides: Experiment and Simulation

Appl Spectrosc. 2019 Dec;73(12):1451-1460. doi: 10.1177/0003702819867342. Epub 2019 Aug 22.

Abstract

The goal of this technical note was to compare experimentally and via simulation of eight substrate-integrated hollow waveguide (iHWG) designs, and to predict promising future iHWG structures in lieu of experiments. The iHWGs differed in their geometry (i.e., inlet funnel cross-section and inner channel cross-section), as well as in their material properties (i.e., type of metal, polish of inner channel). Experimentally, calibration functions of isobutane as a model analyte were determined, and the analytical figures of merit, i.e., signal-to-noise ratio, limit of detection, were evaluated for each iHWG. Evaluation of the amount of radiation incident at the real-world and simulated detector revealed that experiment and simulation were in excellent agreement. While material and quality of the inner channel wall did not have a significant influence on the performance, the iHWG geometry profoundly affected the performance in terms of light throughput: Increasing the inlet funnel dimensions and the inner channel cross-section benefits light throughout, and thus, the analytical signal. Based on these results, simulations of not yet fabricated iHWGs were performed and promising new iHWG structures were suggested.

Keywords: FRED; FRED optical engineering software; FT-IR; Fourier transform infrared spectroscopy; MIR; Substrate-integrated hollow waveguide; iHWG; infrared sensors; mid-infrared; raytracing simulations.