Technical and clinical performance of the thermo-test device "Q-Sense" to assess small fibre function: A head-to-head comparison with the "Thermal Sensory Analyzer" TSA in diabetic patients and healthy volunteers

Eur J Pain. 2019 Nov;23(10):1863-1878. doi: 10.1002/ejp.1461. Epub 2019 Sep 9.

Abstract

Background: Thermo-test devices are rarely used outside specialized pain centres because of high acquisition costs. Recently, a new, portable device ("Q-Sense") was introduced, which is less expensive but has reduced cooling capacity (20°C). We assessed the reliability/validity of the "Q-Sense" by comparing it with the Thermal Sensory Analyzer (TSA).

Methods: Using a phantom-skin model, the physical characteristics of both devices were compared. The clinical performance was assessed in a multicentre study by performing Quantitative Sensory Testing (QST) in 121 healthy volunteers and 83 diabetic patients (Eudra-Med-No. CIV-12-05-006501).

Results: Both device types showed ~40% slower temperature ramps for heating/cooling than nominal data. Cold/warm detection thresholds (CDT, WDT) and heat pain thresholds (HPT) of healthy subjects did not differ between device types. Cold pain thresholds (CPT) were biased for Q-Sense by a floor effect (p < .001). According to intraclass correlation coefficients (ICC), agreement between TSA and Q-Sense was good/excellent for CDT (ICC = 0.894) and WDT (ICC = 0.898), moderate for HPT (ICC = 0.525) and poor for CPT (ICC = 0.305). In diabetic patients, the sensitivity of Q-Sense to detect cold hypoesthesia was reduced in males >60 years. Moderate correlations between thermal detection thresholds and morphological data from skin biopsies (n = 51) were similar for both devices.

Conclusions: Physical characteristics of both thermo-test devices are similarly limited by the poor temperature conduction of the skin. The Q-Sense is useful for thermal detection thresholds but of limited use for pain thresholds. For full clinical use, the lower cut-off temperature should be set to ≤18°C.

Significance: High purchase costs prevent a widespread use of thermo-test devices for diagnosing small fibre neuropathy. The air-cooled "Q-Sense" could be a lower cost alternative, but its technical/clinical performance needs to be assessed because of its restricted cut-off for cooling (20°C). This study provides critical information on the physical characteristics and the clinical validity/reliability of the Q-Sense compared to the "Thermal Sensory Analyzer" (TSA). We recommend lowering the cut-off value of the Q-Sense to ≤18°C for its full clinical use.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Age Factors
  • Aged
  • Case-Control Studies
  • Cold Temperature*
  • Diabetes Mellitus
  • Diabetic Neuropathies / diagnosis*
  • Diabetic Neuropathies / physiopathology
  • Equipment and Supplies*
  • Female
  • Healthy Volunteers
  • Hot Temperature*
  • Humans
  • Hypesthesia / diagnosis*
  • Hypesthesia / physiopathology
  • Male
  • Middle Aged
  • Pain
  • Pain Measurement
  • Pain Threshold*
  • Reproducibility of Results
  • Sensory Thresholds
  • Sex Factors
  • Skin / innervation
  • Skin / pathology
  • Small Fiber Neuropathy
  • Thermosensing*
  • Young Adult