Optimization of the Oertli CataRhex 3® phacoemulsification machine

Clin Ophthalmol. 2019 Apr 16:13:633-639. doi: 10.2147/OPTH.S196281. eCollection 2019.

Abstract

Purpose: The aim of this study was to determine optimal bottle height, vacuum, aspiration rate, and power settings of the Oertli CataRhex 3® phacoemulsification machine.

Methods: Porcine lens nuclei were hardened with formalin and cut into 2.0 mm cubes. Lens cubes were emulsified using the easyTip® 2.2 mm at 30°. Fragment removal time (efficiency) and fragment bounces off the tip (chatter) were measured. Settings tested included bottle height of 60, 80, 100 and 120 cm; aspiration rate of 40, 45, and 50 mL/min; vacuum of 400, 500, and 600 mmHg; and power of 50, 60, 70, 80, 90, and 100%.

Results: Efficiency and chatter increased in a linear fashion with increasing vacuum to 600 mmHg (P=0.017, P=0.046, respectively). The most efficient aspiration rate was 50 mL/min, although this finding lacked statistical significance (P=0.66). Increasing power increased efficiency up to 80% without increasing chatter (P=0.042, P=0.71, respectively). Compared to all other power settings, chatter was increased at 100% (P=0.014).

Conclusion: The most efficient machine settings were vacuum at 600 mmHg, aspiration rate at 50 mL/min, and power at 80%.

Keywords: aspiration rate; blindness; cataract; power; underserved countries; vacuum.