Mechanical antithrombogenic properties by vibrational excitation of the impeller in a magnetically levitated centrifugal blood pump

Artif Organs. 2019 Sep;43(9):849-859. doi: 10.1111/aor.13541. Epub 2019 Aug 22.

Abstract

Mechanical circulatory support devices have been used clinically for patients with heart failure for over 10 years. However, thrombus formation inside blood pumps remains a risk to patient life, causing pump failure and contributing to neurological damage through embolization. In this article, we propose a method for preventing thrombus formation by applying vibrational excitation to the impeller. We evaluate the ability of this method to enhance the antithrombogenic properties of a magnetically levitated centrifugal blood pump and ensure that the impeller vibration does not cause undue hemolysis. First, 3 vibrational conditions were compared using an isolated pump without a mock circulation loop; the vibrational excitation frequencies and amplitudes for the impeller were set to (a) 0 Hz-0 μm, (b) 70 Hz-10 μm, and (c) 300 Hz-2.5 μm. The motor torque was measured to detect thrombus formation and obtain blood coagulation time by calculating the derivative of the torque. Upon thrombus detection, the pump was stopped and thrombi size were evaluated. The results showed an increase in the blood coagulation time and a decrease in the rate of thrombus formation in pumps with the impeller vibration. Second, an in vitro hemolysis test was performed for each vibrational condition to determine the effect of impeller vibration on hemolysis. The results revealed that there was no significant difference in hemolysis levels between each condition. Finally, the selected vibration based on the above test results and the non-vibration as control were compared to investigate antithrombogenic properties under the continuous flow condition. The blood coagulation time and thrombi size were investigated. As a result, vibrational excitation of the impeller at a frequency of 300 Hz and amplitude of 2.5 μm was found to significantly lengthen clotting time, decreasing the rate of pump thrombus compared to the non-vibration condition. We indicate the potential of impeller vibration as a novel mechanical antithrombogenic mechanism for rotary blood pumps.

Keywords: antithrombogenicity; hemolysis; intelligent function; magnetic bearing; vibrational excitation of impeller.

MeSH terms

  • Animals
  • Assisted Circulation / adverse effects
  • Assisted Circulation / instrumentation*
  • Blood Coagulation
  • Centrifugation / adverse effects
  • Centrifugation / instrumentation
  • Equipment Design
  • Heart-Assist Devices / adverse effects*
  • Hemolysis
  • Humans
  • Magnetics / instrumentation
  • Swine
  • Thrombosis / blood
  • Thrombosis / etiology*
  • Thrombosis / prevention & control
  • Vibration