Background: Humidification is a standard of care during invasive mechanical ventilation. Two types of devices are used for this purpose: heated humidifiers and heat-and-moisture exchangers (HME).
Aim: To compare the short-term physiologic effects of an active HME, with those of heated humidifiers and HMEs in terms of respiratory effort, ventilatory pattern, and arterial blood gases during invasive mechanical ventilation.
Methods: We conducted a randomized crossover study with 3 different devices in 15 stable subjects who had a tracheostomy and were ventilator-dependent. Transdiaphragmatic pressure, ventilatory pattern, arterial blood gases, and dyspnea scale were recorded at baseline and at the end of a 20-min period with each device.
Results: Compared with heated humidifiers, the active HME was associated with higher diaphragmatic pressure-time product per minute (117.10 [interquartile range {IQR} 34.58-298.60]) versus 80.86 (IQR, 25.46-110.55) cm H2O×s/min, P = .01), higher PaCO2 (48.50 [IQR, 40.65-53.70] vs 39.60 [IQR, 37.50-49.95]) mm Hg, P = .02) and lower pH (7.41 [IQR, 7.36-7.49] vs 7.45 [IQR, 7.40-7.51], P = .030) without any significant difference in ventilatory pattern. A significantly worse dyspnea scale score (active HME, 3 (2-4) vs heated humidifiers: 4 (3-5); P = .009) was also observed. No significant differences were seen between active HME and HME.
Conclusions: This study indicated that, compared with the heated humidifiers, the use of the active HME or the HME increased inspiratory effort, PaCO2 , pH, and dyspnea in stable subjects who were tracheostomized and ventilator-dependent. (ClinicalTrials.gov registration NCT02499796.).
Keywords: critical care; critically ill subjects; humidifiers; prolonged mechanical ventilation; tracheostomy; work of breathing.