Body Fluid Testing at John F. Kennedy Medical Center in Liberia

Am J Clin Pathol. 2019 Jun 5;152(1):86-90. doi: 10.1093/ajcp/aqz027.

Abstract

Objectives: To apply a simple method to validate testing for albumin, glucose, lactate dehydrogenase (LDH) and total protein (TP) in peritoneal, pleural, and cerebrospinal fluids (CSF) at a hospital in Liberia.

Methods: Serum and body fluid specimens were mixed to create 100% serum and 25%, 50%, 75%, and 100% fluid tubes, which were tested on a Biotecnica BT3500. Differences less than 10% between calculated and measured concentrations were considered acceptable.

Results: The means (confidence intervals) of the percent differences were: albumin/peritoneal 12.8 (6.0-19.7), albumin/pleural 2.8 (1.3-4.2), albumin/CSF 4.8 (2.2-7.5), glucose/peritoneal 4.0 (1.9-6.0), glucose/pleural 4.4 (3.1-5.7), glucose/CSF 2.9 (1.8-4.0), LDH/peritoneal 9.5 (6.3-12.7), LDH/pleural 9.5 (5.4-13.6), LDH/CSF 9.2 (5.2-13.3), TP/peritoneal 7.6 (3.8-11.4), TP/pleural 3.8 (1.5-6.2), and TP/CSF 4.5 (1.0-8.1).

Conclusions: All mean differences except for one were less than 10%, allowing for the adoption of clinical testing. The mixing study is a low-cost method for quality-assured testing that can be performed by resource-limited laboratories.

Keywords: Body fluid; Clinical chemistry; Global health; Liberia; Quality improvement; Validation.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study