Edge-to-Edge Mitral Valve Repair With Extended Clip Arms: Early Experience From a Multicenter Observational Study

JACC Cardiovasc Interv. 2019 Jul 22;12(14):1356-1365. doi: 10.1016/j.jcin.2019.03.023. Epub 2019 May 22.

Abstract

Objectives: The aim of this study was to investigate the technical success and efficacy of mitral valve edge-to-edge repair using extended clip arms.

Background: A new iteration of the MitraClip system, the MitraClip XTR, was introduced in 2018 with the aim of addressing technical limitations observed with previous versions.

Methods: Patients having received at least 1 new implant for the treatment of symptomatic mitral regurgitation (MR) were eligible for this study.

Results: Among the 107 patients (mean age 76 ± 9 years, 69% men) included in this study, the etiology of MR was balanced, with one-half (n = 53 [50%]) classified as secondary and the remaining 54 patients having either primary (n = 40 [37%]) or mixed (n = 14 [13%]) disease. The mean number of devices implanted was 1.5 ± 0.6. Multiple device implantation was required in 46 patients (43%). Single-leaflet device attachment occurred in 4 patients and leaflet injury in 2 additional patients, requiring surgical conversion in 4 patients. Among the 102 patients discharged alive without mitral valve surgery, 95 (93%) had MR ≤2+ and 79 (77%) had MR ≤1+. The mean transmitral gradient increased from 1.9 ± 1.0 mm Hg at baseline to 3.5 ± 1.8 mm Hg at discharge (p < 0.001).

Conclusions: Technical success with the new mitral valve repair system with extended clip arm was achieved in 93% of the patients. MR ≤2+ was obtained in 95 patients (93%) and MR ≤1+ in 79 (77%). The main reasons for procedural failure were acute single-leaflet device attachment associated with leaflet damage or isolated leaflet injury and often required surgical correction.

Keywords: MitraClip; mitral regurgitation; mitral valve; transcatheter mitral valve repair.

Publication types

  • Multicenter Study
  • Observational Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheterization / mortality
  • Female
  • Germany
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Mitral Valve / diagnostic imaging
  • Mitral Valve / physiopathology
  • Mitral Valve / surgery*
  • Mitral Valve Insufficiency / diagnostic imaging
  • Mitral Valve Insufficiency / mortality
  • Mitral Valve Insufficiency / physiopathology
  • Mitral Valve Insufficiency / surgery*
  • Prosthesis Design
  • Recovery of Function
  • Risk Factors
  • Switzerland
  • Time Factors
  • Treatment Outcome