Purpose: To investigate the feasibility of single-needle high-frequency irreversible electroporation (SN-HFIRE) to create reproducible tissue ablations in an in vivo pancreatic swine model.
Materials and methods: SN-HFIRE was performed in swine pancreas in vivo in the absence of intraoperative paralytics or cardiac synchronization using 3 different voltage waveforms (1-5-1, 2-5-2, and 5-5-5 [on-off-on times (μs)], n = 6/setting) with a total energized time of 100 μs per burst. At necropsy, ablation size/shape was determined. Immunohistochemistry was performed to quantify apoptosis using an anticleaved caspase-3 antibody. A numerical model was developed to determine lethal thresholds for each waveform in pancreas.
Results: Mean tissue ablation time was 5.0 ± 0.2 minutes, and no cardiac abnormalities or muscle twitch was detected. Mean ablation area significantly increased with increasing pulse width (41.0 ± 5.1 mm2 [range 32-66 mm2] vs 44 ± 2.1 mm2 [range 38-56 mm2] vs 85.0 ± 7.0 mm2 [range 63-155 mm2]; 1-5-1, 2-5-2, 5-5-5, respectively; p < 0.0002 5-5-5 vs 1-5-1 and 2-5-2). The majority of the ablation zone did not stain positive for cleaved caspase-3 (6.1 ± 2.8% [range 1.8-9.1%], 8.8 ± 1.3% [range 5.5-14.0%], and 11.0 ± 1.4% [range 7.1-14.2%] cleaved caspase-3 positive 1-5-1, 2-5-2, 5-5-5, respectively), with significantly more positive staining at the 5-5-5 pulse setting compared with 1-5-1 (p < 0.03). Numerical modeling determined a lethal threshold of 1114 ± 123 V/cm (1-5-1 waveform), 1039 ± 103 V/cm (2-5-2 waveform), and 693 ± 81 V/cm (5-5-5 waveform).
Conclusions: SN-HFIRE induces rapid, predictable ablations in pancreatic tissue in vivo without the need for intraoperative paralytics or cardiac synchronization.
Copyright © 2019 SIR. Published by Elsevier Inc. All rights reserved.