Development of a rotational set-up correction device for stereotactic head radiation therapy: A performance evaluation

J Appl Clin Med Phys. 2019 Jun;20(6):206-212. doi: 10.1002/acm2.12616. Epub 2019 May 21.

Abstract

We developed a new head supporting device to provide accurate correction of rotational setup during image-guided radiation therapy (IGRT), evaluating its correction performance and the efficacy of dose distribution in stereotactic radiotherapy (SRT) using a helical tomotherapy (HT) system. The accuracy of rotational motion was measured using an electronic inclinometer; we compared device angles and measurement values from 0.0° to 3.0°. The correction accuracy was investigated based on the distance between rotational centers in the device and on megavoltage computed tomography (MVCT); the correction values were compared using distances in the range of 0.0-9.0 cm using a head phantom with a rotational error of 1.5°. For an SRT with a simultaneous integrated boost plan and a rotational error of 3.0° in yaw angle using a head phantom, and for a single-isocenter SRT for multiple brain metastases in the data of three patients, dosimetric efficacy of the HT unit was evaluated for calculated dose distributions with MVCT after rotational correction. This device can correct pitch and yaw angles within 0.3° and can be corrected to within 0.5° for each rotational angle according to the result of MVCT correction regardless of the rotational center position. In the head phantom study, the device had a beneficial impact on rotational correction; D99% for the target improved by approximately 10% with rotational correction. Using patient data with the device, the mean difference based on the treatment planning data was 0.3% for D99% and -0.1% for coverage index to the target. Our rotational setup correction device has high efficacy, and can be used for IGRT.

Keywords: head supporting device; helical tomotherapy; multiple targets; stereotactic radiotherapy.

Publication types

  • Evaluation Study

MeSH terms

  • Brain Neoplasms / surgery*
  • Head / radiation effects*
  • Humans
  • Organs at Risk / radiation effects
  • Patient Positioning*
  • Phantoms, Imaging*
  • Radiosurgery / methods*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy Setup Errors / prevention & control*
  • Radiotherapy, Intensity-Modulated / methods
  • Retrospective Studies
  • Rotation
  • Surgery, Computer-Assisted / methods*