Validation of Smartphone-Based Retinal Photography for Diabetic Retinopathy Screening

Ophthalmic Surg Lasers Imaging Retina. 2019 May 1;50(5):S18-S22. doi: 10.3928/23258160-20190108-05.

Abstract

Backround and objective: Screening for diabetic retinopathy (DR) is cost-effective when compared with disability loss for those who go blind in the absence of a screening program. We aimed to evaluate the sensitivity and specificity of a smartphone-based device for the screening and detection of DR.

Patients and methods: A cross-sectional study of 220 patients with diabetes (440 eyes, all patients age 25 years or older) was completed. Tropicamide 0.5% was used for iris dilation followed by an indirect ophthalmoscopy using a 20-D lens. Retinal images were later obtained using a smartphone attached to an adaptable camera device. Retinal images permitted the visualization of the macular and papillary regions and were sent without compression via the internet to a retinal specialist for interpretation. Sensitivity and specificity were calculated for all cases and stages of DR.

Results: Using our standard examination method, the prevalence of DR and macular edema were 13.6% and 6.4%, respectively. With the smartphone-based retinal camera, the prevalence of DR and macular edema were 18.2% and 8.2%, respectively. Sensitivity and specificity for the detection of all stages of DR was 73.3% and 90.5%, respectively. For the detection of macular edema, sensitivity was 77.8%, and specificity was 95%. For severe nonproliferative DR (NPDR), sensitivity and specificity were 80% and 99%, respectively; for proliferative DR (PDR), they were both 100%. In the early stages of DR, specificity was 89.8% for mild NPDR and 97.1% for moderate NPDR. Sensitivity was 57.1% and 42.9%, respectively.

Conclusion: Screening for DR using a smartphone-based retinal camera has a satisfactory specificity at all DR stages. Its sensitivity seems to be high only in the stages of DR necessitating a specific therapeutic decision (eg, macular edema, severe NPDR, and PDR). A smartphone-based retinal camera may be a useful device to screen for DR in resource-limited settings. [Ophthalmic Surg Lasers Imaging Retina. 2019;50:S18-S22.].

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Cross-Sectional Studies
  • Diabetic Retinopathy / diagnosis*
  • Equipment Design
  • Female
  • Humans
  • Male
  • Mass Screening / methods*
  • Middle Aged
  • Photography / instrumentation*
  • Reproducibility of Results
  • Retina / diagnostic imaging*
  • Retrospective Studies
  • Smartphone*