Upper reference limits of transient ischemic dilation ratio for different protocols on new-generation cadmium zinc telluride cameras: A report from REFINE SPECT registry

J Nucl Cardiol. 2020 Aug;27(4):1180-1189. doi: 10.1007/s12350-019-01730-y. Epub 2019 May 13.

Abstract

Background: Upper reference limits for transient ischemic dilation (TID) have not been rigorously established for cadmium-zinc-telluride (CZT) camera systems. We aimed to derive TID limits for common myocardial perfusion imaging protocols utilizing a large, multicenter registry (REFINE SPECT).

Methods: One thousand six hundred and seventy-two patients with low likelihood of coronary artery disease with normal perfusion findings were identified. Images were processed with Quantitative Perfusion SPECT software (Cedars-Sinai Medical Center, Los Angeles, CA). Non-attenuation-corrected, camera-, radiotracer-, and stress protocol-specific TID limits in supine position were derived from 97.5th percentile and mean + 2 standard deviations (SD). Reference limits were compared for different solid-state cameras (D-SPECT vs. Discovery), radiotracers (technetium-99m-sestamibi vs. tetrofosmin), different types of stress (exercise vs. four different vasodilator-based protocols), and different vasodilator-based protocols.

Results: TID measurements did not follow Gaussian distribution in six out of eight subgroups. TID limits ranged from 1.18 to 1.52 (97.5th percentile) and 1.18 to 1.39 (mean + 2SD). No difference was noted between D-SPECT and Discovery cameras (P = 0.71) while differences between exercise and vasodilator-based protocols (adenosine, regadenoson, or regadenoson-walk) were noted (all P < 0.05).

Conclusions: We used a multicenter registry to establish camera-, radiotracer-, and protocol-specific upper reference limits of TID for supine position on CZT camera systems. Reference limits did not differ between D-SPECT and Discovery camera.

Keywords: Coronary artery disease; image analysis; myocardial perfusion imaging; perfusion agents; single-photon emission computed tomography; vasodilators.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Cadmium
  • Female
  • Gamma Cameras*
  • Humans
  • Male
  • Middle Aged
  • Myocardial Ischemia / diagnostic imaging*
  • Myocardial Perfusion Imaging / methods*
  • Registries
  • Tellurium
  • Tomography, Emission-Computed, Single-Photon / methods*
  • Zinc

Substances

  • CdZnTe
  • Cadmium
  • Zinc
  • Tellurium