Spatial and temporal pre-alignment of an X-ray split-and-delay unit by laser light interferometry

Rev Sci Instrum. 2019 Apr;90(4):045106. doi: 10.1063/1.5089496.

Abstract

We present a novel experimental setup for performing a precise pre-alignment of a hard X-ray split-and-delay unit based on low coherence light interferometry and high-precision penta-prisms. A split-and-delay unit is a sophisticated perfect crystal-optics device that splits an incoming X-ray pulse into two sub-pulses and generates a controlled time-delay between them. While the availability of a split-and-delay system will make ultrafast time-correlation and X-ray pump-probe experiments possible at free-electron lasers, its alignment process can be very tedious and time-consuming due to its complex construction. By implementing our experimental setup at beamline P10 of PETRA III, we were able to reduce the time of alignment to less than 3 h. We also propose an alternate method for finding the zero-time delay crossing without the use of X-rays or pulsed laser sources. The successful demonstration of this method brings prospect for operating the split-and-delay systems under alignment-time-critical environments such as X-ray free electron laser facilities.