Breakage of metal hip arthroplasty components: Retrieval and structural analysis

Biomed Mater Eng. 2019;30(3):297-308. doi: 10.3233/BME-191053.

Abstract

Background: Breakage of joint arthroplasty components are rare, yet during an implant retrieval program we found several cases.

Objective: In this study we examined the components to determine the causes and mechanisms of breakage of these implants.

Methods: From our collection of 849 retrievals we selected 682 cases with metal parts (503 hip, 79 knee arthroplasties) and identified fractured components: seven hip resurfacing implants, five total hip replacement stems, one monopolar femoral head, and one modular revision femoral stem from. Implants were examined using optical and scanning electron microscopy; metallographic sections were prepared and samples of periprosthetic tissues underwent microscopic examination.

Results: In the resurfacing components breakage occurred in small stems placed in the femoral neck due to necrosis of femoral heads, with no metal flaws detected. Fatigue breakage of femoral stems was caused by presence of material flaws in the CoCrMo alloy, and corrosion. The monopolar head failed in fatigue fracture mechanism, breakage was initiated in an undercut near the taper connection for femoral component. The modular stem from Ti alloy sustained fatigue fracture induced by corrosion caused by debris from previously revised stem; no material flaws were detected in this sample. In most cases periprosthetic tissues had a morphology typical for aseptic loosening.

Conclusions: In our series failure was caused by material flaws, presence of stress raisers and localized corrosion. Our findings indicate that sharp edges and other features which can act as stress raisers should be avoided in newly designed implants. Corrosion induced fracture of the modular Ti stem indicates the need for a detailed debridement of periprosthetic tissues during revision arthroplasties.

Keywords: Retrieval analysis; component breakage; hip arthroplasty.

MeSH terms

  • Arthroplasty, Replacement, Hip / adverse effects
  • Corrosion
  • Equipment Failure Analysis
  • Femur Head / pathology
  • Femur Head / surgery
  • Hip Joint / surgery
  • Hip Prosthesis* / adverse effects
  • Humans
  • Prosthesis Design
  • Prosthesis Failure* / adverse effects
  • Vitallium* / adverse effects
  • Vitallium* / chemistry

Substances

  • Vitallium