Evaluating the utility of elemental measurements obtained from factory-calibrated field-portable X-Ray fluorescence units for aquatic sediments

Environ Pollut. 2019 Jun:249:45-53. doi: 10.1016/j.envpol.2019.03.001. Epub 2019 Mar 5.

Abstract

We assessed factory-calibrated field-portable X-ray fluorescence (pXRF) data quality for use with minimally-prepared aquatic sediments, including the precision of replicate pXRF measurements, accuracy of factory-calibrated pXRF values as compared to total digestion/ICP-OES concentrations, and comparability of calibrated pXRF values to extractable concentrations. Data quality levels for precision, accuracy, and comparability were not equivalent for element/analyzer combinations. All analyses of elements that were assessed for precision and accuracy on a single analyzer were both precise (<10% relative standard deviation) and accurate (r2 > 0.85) for K, Ca, Ti, Mn, Fe, and Zn. Calibrated pXRF values for Al, K, Ca, Ti, Mn, Fe, Cu, Zn, and Pb were within ∼10% relative difference of total digestion/ICP-OES concentrations. Calibrated pXRF values for Fe, Cu, Zn, As, and Pb were within ∼20% relative difference of extractable concentrations. Some elements had a higher level of data quality using specific analyzers, but in general, no pXRF analyzer had the highest level of data quality in all categories. Collectively, our data indicate that a wide range of factory-calibrated pXRF units are capable of providing high-quality total concentrations for the analysis of aquatic sediments.

MeSH terms

  • Calibration
  • Environmental Monitoring / instrumentation
  • Environmental Monitoring / methods*
  • Geologic Sediments / chemistry
  • Spectrometry, X-Ray Emission*
  • Water Pollutants / analysis*

Substances

  • Water Pollutants