Purpose: B0 field inhomogeneity may cause significant errors in chemical shift encoding-based fat-water (F/W) separation. We describe a new approach to improve its robustness using novel B0 field map pre-estimation.
Methods: Our method exploits insensitivity of fat to magnetization transfer effect, which allows generating fat-insensitive B0 field priors with full or partial spatial support using a low-resolution magnetization transfer-weighted scan. The full prior can be employed by most F/W separation methods for initialization or data demodulation. We also propose a modified region-growing algorithm in which the partial prior is utilized for its initial seeding.
Results: The magnetization transfer-based B0 priors significantly reduced F/W errors of three representative F/W separation methods in all cases. In cases with moderate B0 inhomogeneity, the full prior allowed error-free separation even with basic, voxel-independent processing. When coupled with methods exploiting B0 field smoothness, it significantly improved separation accuracy even in the presence of strong inhomogeneities. Seeding the region-growing with the partial prior significantly improved performance of F/W separation, including cases with spatially disconnected tissues.
Conclusion: Magnetization transfer-based B0 field pre-estimation provides valuable prior information for F/W separation, which may significantly improve its robustness at the expense of nominal (< 5%-10%) scan time increase.
Keywords: Chemical shift imaging; Dixon fat-water separation; IDEAL; magnetization transfer.
© 2019 International Society for Magnetic Resonance in Medicine.