Belt-Mounted Micro-Gas-Chromatograph Prototype for Determining Personal Exposures to Volatile-Organic-Compound Mixture Components

Anal Chem. 2019 Apr 2;91(7):4747-4754. doi: 10.1021/acs.analchem.9b00263. Epub 2019 Mar 19.

Abstract

We describe a belt-mountable prototype instrument containing a gas chromatographic microsystem (μGC) and demonstrate its capability for near-real-time recognition and quantification of volatile organic compounds (VOCs) in moderately complex mixtures at concentrations encountered in industrial workplace environments. The μGC comprises three discrete, Si/Pyrex microfabricated chips: a dual-adsorbent micropreconcentrator-focuser for VOC capture and injection; a wall-coated microcolumn with thin-metal heaters and temperature sensors for temperature-programmed separations; and an array of four microchemiresistors with thiolate-monolayer-protected-Au-nanoparticle interface films for detection and recognition-discrimination. The battery-powered μGC prototype (20 × 15 × 9 cm, ∼2.1 kg sans battery) has on-board microcontrollers and can autonomously analyze the components of a given VOC mixture several times per hour. Calibration curves bracketing the Threshold Limit Value (TLV) of each VOC yielded detection limits of 16-600 parts-per-billion for air samples of 5-10 mL, well below respective TLVs. A 2:1 injection split improved the resolution of early eluting compounds by up to 63%. Responses and response patterns were stable for 5 days. Use of retention-time windows facilitated the chemometric recognition and discrimination of the components of a 21-VOC mixture sampled and analyzed in 3.5 min. Results from a "mock" field test, in which personal exposures to time-varying concentrations of a mixture of five VOCs were measured autonomously, agreed closely with those from a reference GC. Thus, reliable, near-real-time determinations of worker exposures to multiple VOCs with this wearable μGC prototype appear feasible.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Air Pollution, Indoor / analysis*
  • Breath Tests* / instrumentation
  • Chromatography, Gas
  • Environmental Monitoring* / instrumentation
  • Humans
  • Volatile Organic Compounds / administration & dosage
  • Volatile Organic Compounds / analysis*

Substances

  • Volatile Organic Compounds