Faecal calprotectin determination: impact of preanalytical sample treatment and stool consistency on within- and between-method variability

Biochem Med (Zagreb). 2019 Feb 15;29(1):010707. doi: 10.11613/BM.2019.010707.

Abstract

Introduction: We assessed the differences in faecal calprotectin (FC) concentrations measured by two assays depending on the stool consistency and extraction method.

Materials and methods: Stool samples were extracted using the EliA Stool Extraction Kit, Calex® Cap extraction device and respective weighing methods, while FC concentrations were measured using the EliATM Calprotectin and Bühlmann fCAL® Turbo method and checked for within- and between-method variability with regard to extraction method and stool consistency category. Extraction yield was evaluated for impact of different sample incubation time (10 min and 1 h) in extraction buffer for both methods and for impact of different initial sample dilutions (1:50, 1:100, 1:500) for fCAL® Turbo method.

Results: Results determined from Calex® Cap extracts were higher compared to weighing method extracts (mean bias 33.3%; P < 0.001), while no significant difference was found between results obtained with EliA Stool Extraction Kit and weighing method (mean bias 0.1%; P = 0.484), in both cases irrespective of stool consistency. Bühlmann fCAL® Turbo results were higher than EliATM Calprotectin results (mean bias 32.3%, P = 0.025 weighing method; and mean bias 53.9%, P < 0.001 extraction devices), the difference is dependent on stool consistency and FC concentration. Significantly higher FC extraction yield was obtained with longer sample incubation time for both methods (P = 0.019 EliATM Calprotectin; P < 0.001 fCAL® Turbo) and with increasing initial sample dilution for fCAL® Turbo method (P < 0.001).

Conclusion: Preanalytical stool sample handling proved to be a crucial factor contributing to within- and between-FC assay variability. Standardization is urgently needed in order to assure comparable and reliable FC results.

Keywords: calprotectin; immunoassay; inflammatory bowel disease; preanalytical phase.

MeSH terms

  • Feces / chemistry*
  • Humans
  • Leukocyte L1 Antigen Complex / analysis*
  • Specimen Handling*

Substances

  • Leukocyte L1 Antigen Complex