A linear programming approach to inverse planning in Gamma Knife radiosurgery

Med Phys. 2019 Apr;46(4):1533-1544. doi: 10.1002/mp.13440. Epub 2019 Mar 8.

Abstract

Purpose: Leksell Gamma Knife® is a stereotactic radiosurgery system that allows fine-grained control of the delivered dose distribution. We describe a new inverse planning approach that both resolves shortcomings of earlier approaches and unlocks new capabilities.

Methods: We fix the isocenter positions and perform sector-duration optimization using linear programming, and study the effect of beam-on time penalization on the trade-off between beam-on time and plan quality. We also describe two techniques that reduce the problem size and thus further reduce the solution time: dualization and representative subsampling.

Results: The beam-on time penalization reduces the beam-on time by a factor 2-3 compared with the naïve alternative. Dualization and representative subsampling each leads to optimization time-savings by a factor 5-20. Overall, we find in a comparison with 75 clinical plans that we can always find plans with similar coverage and better selectivity and beam-on time. In 44 of these, we can even find a plan that also has better gradient index. On a standard GammaPlan workstation, the optimization times ranged from 2.3 to 26 s with a median time of 5.7 s.

Conclusion: We present a combination of techniques that enables sector-duration optimization in a clinically feasible time frame.

Keywords: Leksell Gamma Knife®; inverse planning; optimization; stereotactic radiosurgery; treatment planning.

MeSH terms

  • Algorithms
  • Cavernous Sinus / pathology
  • Cavernous Sinus / surgery*
  • Humans
  • Meningeal Neoplasms / pathology
  • Meningeal Neoplasms / surgery*
  • Meningioma / pathology
  • Meningioma / surgery*
  • Programming, Linear*
  • Radiosurgery / methods
  • Radiosurgery / standards*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / standards*