Comparison of Skull Radiograph and Computed Tomography Measurements of Cochlear Implant Insertion Angles

Otol Neurotol. 2019 Mar;40(3):e298-e303. doi: 10.1097/MAO.0000000000002121.

Abstract

Background: Measurement of the angular depth of insertion (aDOI) of cochlear implant electrode arrays has numerous clinical and research applications. Plain-film radiographs are easily obtained intraoperatively and have been described as a means to calculate aDOI. CT imaging with 3D reformatting can also be used for this measurement, but is less conveniently obtained and requires higher radiation doses, a particular concern in pediatrics. The extent to which plain-film and 3D CT image-based measurements are representative of the true position of the electrode within the cochlea is unknown.

Methods: Cochlear implantation was performed on 10 cadaveric temporal bones. Five bones were implanted with perimodiolar electrodes (Contour Advance TM, Cochlear, Sydney, Australia) and five were implanted with lateral wall electrodes (Slim Straight, Cochlear). The insertion depths of the electrodes were varied. Each bone was imaged with a radiograph and CT. aDOI was measured for each bone in each imaging modality by a neurotologist and a neuroradiologist. To obtain a 'gold standard' estimate of aDOI, the implanted temporal bones were embedded in an epoxy resin and methodically sectioned at 100 μm intervals; histologic images were captured at each interval. A 3D stack of the images was compounded, and a MATLAB script used to calculate aDOI of the most apical electrode. Measurements in the three modalities (radiograph, CT, and histology) were then compared.

Results: The average aDOI across all bones was similar for all modalities: 423° for radiographs, 425° for CT scans, and 427° for histology, indicating that neither imaging modality resulted in large systematic errors. Using the histology-measured angles as a reference, the average error for CT-based measures (regardless of whether the error was in the positive or negative direction) was 12°, and that for radiograph-based measures was 15°. This small difference (12 vs 15° error) was not statistically significant.

Conclusion: Based on this cadaveric temporal bone model, both radiographs and CTs can provide reasonably accurate aDOI measurements. In this small sample, and as expected, the CT-based estimates were more accurate than the radiograph-based measurements. However, the difference was small and not statistically significant. Thus, the use of plain radiographs to calculate aDOI seems judicious whenever it is desired to prevent unnecessary radiation exposure and expense.

Publication types

  • Comparative Study
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Australia
  • Cadaver
  • Cochlea / diagnostic imaging*
  • Cochlea / surgery
  • Cochlear Implantation* / methods
  • Cochlear Implants
  • Humans
  • Imaging, Three-Dimensional / methods*
  • Neuroimaging / methods*
  • Radiography / methods
  • Temporal Bone / diagnostic imaging*
  • Temporal Bone / surgery
  • Tomography, X-Ray Computed / methods