Susceptibility of Cardiac Troponin Assays to Biotin Interference

Am J Clin Pathol. 2019 Apr 2;151(5):486-493. doi: 10.1093/ajcp/aqy172.

Abstract

Objectives: To investigate biotin interference on three cardiac troponin (cTn) assays and demonstrate a method to overcome biotin interference.

Methods: cTn levels were measured in (1) plasma from healthy volunteers on 10-mg daily biotin supplementation mixed with a plasma with known elevated troponin, (2) plasmas with known elevated cTn after mixing in reagent biotin to simulate supplementation, and (3) biotin-spiked plasma specimens pretreated with streptavidin-agarose beads.

Results: Daily biotin ingestion (10 mg) and studies simulating daily biotin use resulted in significant interference in the Gen5 cardiac troponin T (cTnT) assay; the contemporary Gen 4 cTnT and high-sensitivity cardiac troponin I (hs-cTnI) assays were unaffected. The biotin interference threshold was 31, 315, and more than 2,000 ng/mL for Gen5 cTnT, cTnT, and hs-cTnI assays, respectively. Streptavidin pretreatment blocked biotin interference in cTn assays.

Conclusions: Biotin interference is possible at plasma concentrations achievable by ingestion of over-the-counter supplements that may lead to delayed or missed diagnosis of myocardial injury with the Gen5 cTnT assay.

Keywords: Biotin; Interference; Myocardial infarction; Troponin.