[Validity study of Coda Motion Analysis System for measuring cervical lateral flexion in normal adults]

Zhongguo Gu Shang. 2018 Nov 25;31(11):1017-1021. doi: 10.3969/j.issn.1003-0034.2018.08.007.
[Article in Chinese]

Abstract

Objective: To investigate the validity of Coda Motion Analysis System for measuring cervical lateral flexion in normal adults in order to explore a new measuring tool for clinical and research practice.

Methods: A total of 43 participants were involved in the study. Cervical range of lateral flexion were measured with Coda Motion Analysis System and "gold standard" X-ray simultaneously. The validity and agreement were assessed using the scatter diagram, the Pearson correlation coefficient and limits of agreement.

Results: Cervical range of lateral flexion measured by Coda Motion Analysis System had no statistical differences with those measured by X-ray(P>0.05). The Coda Motion Analysis System demonstrated a very good linear relation with the X-ray measurements in cervical range of right lateral flexion, left flexion and total lateral flexion, and the Pearson correlation coefficients were 0.72, 0.85 and 0.90 respectively.

Conclusions: Coda Motion Analysis System showed good validity for measuring cervical lateral flexion in normal adults. Because the reliability of Coda Motion Analysis System was established previously, the results of this study suggest that the system has the potential to be used to measure cervical lateral flexion in clinical and research practice.

Keywords: Cervical range of motion; Coda Motion Analysis System; Cortical bone trajectory; Lateral flexion.

MeSH terms

  • Adult
  • Cervical Vertebrae*
  • Eye Diseases, Hereditary*
  • Humans
  • Neck
  • Range of Motion, Articular
  • Reproducibility of Results