Optimization of Sample Preparation for Detection of 10 Phthalates in Non-Alcoholic Beverages in Northern Vietnam

Toxics. 2018 Nov 19;6(4):69. doi: 10.3390/toxics6040069.

Abstract

A novel method was developed for the sensitive, cheap and fast quantitation of 10 phthalates in non-alcoholic beverages by liquid⁻liquid extraction (LLE) combined with gas chromatography tandem mass spectrometry (GC-MS/MS). The best results were obtained when n-hexane was used as extraction solvent. A central composite design (CCD) was applied to select the most appreciated operating condition. The method performance was evaluated according to the SANTE/11945/2015 guidelines and was linear in the 0.1 to 200 µg/L range for 10 phthalate compounds, with r² > 0.996 and individual residuals <15%. Repeatability (RSDr), within-laboratory reproducibility (RSDwr), and the trueness range were from 2.7 to 9.1%, from 3.4 to 14.3% and from 91.5 to 118.1%, respectively. The limit of detection (LOD) was between 0.5 to 1.0 ng/L and the limit of quantitation (LOQ) was between 1.5 to 3.0 ng/L for all 10 compounds. The developed method was successfully applied to the analysis of non-alcoholic beverages.

Keywords: GC-MS/MS; liquid–liquid extraction; non-alcoholic beverages; phthalate; response surface methodology.