Effect of sample processing and time delay on cell count and chemistry tests in cerebrospinal fluid collected from drainage systems

Biochem Med (Zagreb). 2018 Oct 15;28(3):030705. doi: 10.11613/BM.2018.030705.

Abstract

Introduction: Cerebrospinal fluid (CSF) from extra-ventricular drainage (EVD) systems is routinely analysed to diagnose EVD-related bacterial meningitis. We investigated the effect of time delay and sample processing on cell count and basic biochemistry results in EVD CSF to define optimal turnaround time and whether manual and automated cell counting are comparable in such samples.

Materials and methods: In total, 32 EVD CSF samples were analysed. Baseline testing included cell counting (Fuchs-Rosenthal chamber and Sysmex XE5000) and biochemistry analyses (glucose, lactate, proteins). Manual cell counting was also performed at intervals of 61-90 and 91-150 minutes from baseline in the residual sample. Biochemistry analyses were performed in samples before and after centrifugation at baseline and at 91-150 minutes interval.

Results: At 91-150 minutes total cell count (P < 0.001), large lymphocytes (P = 0.007), neutrophils (P < 0.001) and phagocytes (P = 0.006) obtained by manual counting decreased and the number of disintegrated cells count increased (P = 0.016) compared to the baseline values. Considering method comparison, proportional difference between methods for all cell (sub)groups was obtained, whereas polymorphonuclears also showed the constant difference (y = 11.21 + 1.22x). Compared to centrifuged CSF, lower concentration of glucose and lactates were obtained in uncentrifuged samples (P < 0.001) at baseline.

Conclusions: Manual cell counting should be performed within 60 minutes as any delay can alter results. The same counting technique should be used to obtain longitudinally assessable results. Biochemistry tests are stable in uncentrifuged CSF up to 2.5 hours.

Keywords: biochemistry analyses; cell count; cerebrospinal fluid; extra-ventricular drainage system; turnaround time.