VyCAP's puncher technology for single cell identification, isolation, and analysis

Cytometry A. 2018 Dec;93(12):1255-1259. doi: 10.1002/cyto.a.23631. Epub 2018 Nov 14.

Abstract

Here we present the Puncher technology for the isolation of single cells. This technology combines a silicon chip with microwells, fluorescence imaging, and a punching method to isolate and transfer the single cells to standard reaction tubes. The technology is compatible with commercially available downstream workflows and instrumentation. Here we focus on the isolation of CTC but the Puncher technology can be applied to isolate single cells from liquid biopsies and more general from cell suspensions. It is especially suited for cell suspensions that contain: Cells of interest at a frequency of 1 per 10,000 or less A low total number of cells ranging from 1 to 100,000, that are present in a volume of 0.01 to 50 mL. The frequency of appearance of CTC in blood is in the order of the 1 per 106 leukocytes. To be able to isolate the single CTC with the Puncher technology, enrichment of the CTC by a 3 logs reduction of the leukocytes is required. Here we describe the use of Rosettesep and Parsortix as examples of pre-enrichment methods that are compatible with the Puncher technology and further downstream applications. © 2018 International Society for Advancement of Cytometry.

Keywords: Puncher; circulating tumor cells; clonal expansion; microwells; single cell isolation.

MeSH terms

  • Cell Line, Tumor
  • Cell Separation / methods*
  • Humans
  • Leukocytes / pathology
  • Liquid Biopsy / methods
  • Microfluidic Analytical Techniques / methods
  • Neoplastic Cells, Circulating / pathology
  • Single-Cell Analysis / methods*