Clinical Interpretation of Peripheral Pulse Oximeters Labeled "Not for Medical Use"

Ann Fam Med. 2018 Nov;16(6):552-554. doi: 10.1370/afm.2317.

Abstract

The purpose of our study was to clarify limitations of off-label use for low cost nonmedical use (NMU) pulse oximeters by primary care providers. These devices are widely marketed over the Internet and in drugstores but are not intended for medical use or reviewed by the Food and Drug Administration (FDA). Our study compared oxygen saturation (SpO2) in patients from 1 medical use (MU) pulse oximeter to 8 NMU pulse oximeters. Measured arterial oxygenation (SaO2) was compared with SpO2 when available. In patients who were normoxic (SpO2 ≥90%), all oximeters exhibited similar readings. This finding suggests that NMU pulse oximeters may be able to rule out hypoxemia in clinical settings.

Keywords: device; oximetry; regulation; safety.

Publication types

  • Observational Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Blood Gas Analysis / methods*
  • Clinical Competence
  • Clinical Decision-Making / methods
  • Cross-Sectional Studies
  • Decision Support Systems, Clinical
  • Female
  • Humans
  • Hypoxia / diagnosis*
  • Male
  • Middle Aged
  • Off-Label Use*
  • Oximetry / methods*
  • Primary Health Care / methods*
  • Young Adult