Towards Ultra Low-Cost Myoactivated Prostheses

Biomed Res Int. 2018 Oct 4:2018:9634184. doi: 10.1155/2018/9634184. eCollection 2018.

Abstract

In developing countries, due to the high cost involved, amputees have limited access to prosthetic limbs. This constitutes a barrier for this people to live a normal life. To break this barrier, we are developing ultra-low-cost closed-loop myoactivated prostheses that are easy to maintain manufacture and that do not require electrodes in contact with the skin to work effectively. In this paper, we present the implementation for a simple but functional hand prosthesis. Our simple design consists of a low-cost embedded microcontroller (Arduino), a wearable stretch sensor (adapted from electroresistive bands normally used for "insulation of gaskets" against EM fields), to detect residual muscle contraction as direct muscle volumetric shifts and a handful of common, not critical electronic components. The physical prosthesis is a 3D printed claw-style two-fingered hand (PLA plastic) directly geared to an inexpensive servomotor. To make our design easier to maintain, the gears and mechanical parts can be crafted from recovered materials. To implement a closed loop, the amount of closure of prosthesis is fed back to the user via a second stretch sensor directly connected to claw under the form of haptic feedback. Our concept design comprised of all the parts has an overall cost below AUD 30 and can be easily scaled up to more complicated devices suitable for other uses, i.e., multiple individual fingers and wrist rotation.

MeSH terms

  • Artificial Limbs*
  • Humans
  • Printing, Three-Dimensional*
  • Prosthesis Design / economics
  • Prosthesis Design / methods*