Nyquist ghost correction of breast diffusion weighted imaging using referenceless methods

Magn Reson Med. 2019 Apr;81(4):2624-2631. doi: 10.1002/mrm.27563. Epub 2018 Nov 2.

Abstract

Purpose: Correction of Nyquist ghosts for single-shot spin-echo EPI using the standard 3-line navigator often fails in breast DWI because of incomplete fat suppression, respiration, and greater B0 inhomogeneity. The purpose of this work is to compare the performance of the 3-line navigator with 4 data-driven methods termed "referenceless methods," including 2 previously proposed in literature, 1 introduced in this work, and finally a combination of all 3, in breast DWI.

Methods: Breast DWI was acquired for 41 patients with SS SE-EPI. Raw data was corrected offline with the standard 3-line navigator and 4 referenceless methods, which modeled the ghost as a linear phase error and minimized 3 unique cost functions as well as the median solution of all 3. Ghost levels were evaluated based on the signal intensity in the background region, defined by a mask auto-generated from a T1 -weighted anatomical image. Ghost intensity measurements were fit to a linear mixed model including ghost correction method and b-value as covariates.

Results: All 4 referenceless methods outperformed the standard 3-line navigator with statistical significance at all 4 b-values tested (b = 0, 100, 600, and 800 s/mm2 ).

Conclusions: Referenceless methods provide a robust way to reduce Nyquist ghosts in breast DWI without the need for any additional calibration scan.

Keywords: DWI; Nyquist ghost; SE-EPI; breast.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Algorithms
  • Artifacts
  • Breast / diagnostic imaging*
  • Calibration
  • Computer Simulation
  • Diffusion Magnetic Resonance Imaging*
  • Echo-Planar Imaging*
  • Female
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Linear Models
  • Normal Distribution
  • Phantoms, Imaging
  • Signal-To-Noise Ratio