Micro-Coil Design Influences the Spatial Extent of Responses to Intracortical Magnetic Stimulation

IEEE Trans Biomed Eng. 2019 Jun;66(6):1680-1694. doi: 10.1109/TBME.2018.2877713. Epub 2018 Oct 23.

Abstract

Objective: Electrical stimulation via cortically implanted electrodes has been proposed to treat a wide range of neurological disorders. Effectiveness has been limited, however, in part due to the inability of conventional electrodes to activate specific types of neurons while avoiding other types. Recent demonstrations that magnetic stimulation from a micro-coil can selectively activate pyramidal neurons (PNs) while avoiding passing axons suggest the possibility that such an approach can overcome some this limitation and here we use computer simulations to explore how the micro-coil design influences the selectivity with which neurons are activated.

Methods: A computational model was developed to compare the selectivity of magnetic stimulation induced by rectangular-, V-, and W-shaped coil designs. The more promising designs (V- and W-shapes) were fabricated for use in electrophysiological experiments including in vitro patch-clamp recording and calcium imaging (GCaMP6f) of mouse brain slices.

Results: Both V- and W-shaped coils reliably activated layer 5 (L5) PNs but V-coils were more effective while W-coils were more selective. Activation thresholds with double-loop coils were approximately one-half those of single-loop coils. Calcium imaging revealed that both V- and W-coils better confine activation than electrodes.

Conclusion: Individual design features can influence both the strength as well as the selectivity of micro-coils and can be accurately predicted by computer simulations.

Significance: Our results show that how coil design influences the response of cortical neurons to stimulation and are an important step toward the development of next-generation cortical prostheses.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Animals
  • Axons / physiology
  • Axons / radiation effects
  • Computer Simulation
  • Equipment Design
  • Mice
  • Mice, Inbred C57BL
  • Neural Prostheses*
  • Transcranial Magnetic Stimulation / instrumentation*
  • Transcranial Magnetic Stimulation / methods
  • Visual Cortex / cytology