Long-Term Outcomes After Transcatheter Aortic Valve-in-Valve Replacement

Circ Cardiovasc Interv. 2018 Sep;11(9):e007038. doi: 10.1161/CIRCINTERVENTIONS.118.007038.

Abstract

Background: Data on long-term outcomes after valve-in-valve (ViV) transcatheter aortic valve replacement (TAVR) are scarce. The objective of this study was to determine the long-term clinical outcomes and structural valve degeneration (SVD) over time in patients undergoing ViV-TAVR.

Methods and results: Consecutive patients undergoing ViV-TAVR in 9 centers between 2009 and 2015 were included. Patients were followed yearly, and clinical and echocardiography data were collected prospectively. SVD was defined as subclinical (increase >10 mm Hg in mean transvalvular gradient+decrease >0.3 cm2 in valve area or new-onset mild or moderate aortic regurgitation) and clinically relevant (increase >20 mm Hg in mean transvalvular gradient+decrease >0.6 cm2 in valve area or new-onset moderate-to-severe aortic regurgitation). A total of 116 patients (mean age, 76±11 years; 64.7% male; mean Society of Thoracic Surgeons score, 8.0±5.1%) were included. Balloon- and self-expandable valves were used in 47.9% and 52.1% of patients, respectively, and 30-day mortality was 6.9%. At a median follow-up of 3 years (range, 2-7 years), 30 patients (25.9%) had died, 20 of them (17.2%) from cardiovascular causes. Average mean transvalvular gradients remained stable up to 5-year follow-up ( P=0.92), but clinically relevant SVD occurred in 3/99 patients (3.0%), and 15/99 patients (15.1%) had subclinical SVD. One patient with SVD had redo ViV-TAVR.

Conclusions: About one-fourth of ViV-TAVR recipients had died after a median follow-up of 3 years. Overall valve hemodynamics remained stable over time and clinically relevant SVD was infrequent, but 1 out of 10 patients exhibited some degree of SVD.

Keywords: bioprosthesis; echocardiography; hemodynamics; standard of care; transcatheter aortic valve replacement.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Insufficiency / diagnostic imaging
  • Aortic Valve Insufficiency / mortality
  • Aortic Valve Insufficiency / physiopathology
  • Aortic Valve Insufficiency / surgery*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Echocardiography, Doppler, Color
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / methods
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Prosthesis Design
  • Prosthesis Failure
  • Recovery of Function
  • Reoperation
  • Retrospective Studies
  • Risk Factors
  • Severity of Illness Index
  • Time Factors
  • Transcatheter Aortic Valve Replacement / adverse effects
  • Transcatheter Aortic Valve Replacement / instrumentation*
  • Transcatheter Aortic Valve Replacement / methods
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome