Validation of a Holographic Display for Quantification of Mitral Annular Dynamics by Three-Dimensional Echocardiography

J Am Soc Echocardiogr. 2019 Feb;32(2):303-316.e4. doi: 10.1016/j.echo.2018.08.010. Epub 2018 Oct 5.

Abstract

Background: Three-dimensional (3D) echocardiography with multiplanar reconstruction (MPR) is used clinically to quantify the mitral annulus. MPR images are, however, presented on a two-dimensional screen, calling into question their accuracy. An alternative to MPR is an autostereoscopic holographic display that enables in-depth visualization of 3D echocardiographic data without the need for special glasses. The aim of this study was to validate an autostereoscopic display using sonomicrometry as a gold standard.

Methods: In 11 anesthetized open-chest pigs, sonomicrometric crystals were placed along the mitral annulus and near the left ventricular apex. High-fidelity catheters measured left atrial and ventricular pressures. Adjustments of pre- and afterload were done by constriction of the inferior vena cava and the ascending aorta, respectively. Three-dimensional epicardial echocardiography was obtained from an apical view and converted to the autostereoscopic display. A 3D virtual semitransparent annular surface (VSAS) was generated to measure commissure width (CW), septal-lateral length, area of the mitral annular surface, nonplanarity angle, and the annular height-to-commissure width ratio in mid-systole and late diastole.

Results: Mitral annular measurements from the 3D VSAS derived from the 3D echocardiographic images and autostereoscopic display correlated well with sonomicrometry over a range of loading conditions: CW length (r = 0.98, P < .00001), septal-lateral length (r = 0.98, P < .00001), annular surface area (r = 0.93, P < .001), nonplanarity angle (r = 0.87, P < .001), and annular height-to-commissure width ratio (r = 0.85, P < .01). The 3D VSAS showed better agreement with the sonomicrometric measurements compared with MPR.

Conclusions: Mitral annular measurements using 3D VSAS correlate well with sonomicrometry over a range of loading conditions and may represent a powerful tool for noninvasive quantification of mitral annular dynamics.

Keywords: 3D echocardiography; Mitral valve; Mitral valve annulus; Sonomicrometry.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Atrial Function, Left / physiology*
  • Diastole
  • Echocardiography, Three-Dimensional / methods*
  • Echocardiography, Transesophageal / methods*
  • Female
  • Holography / methods*
  • Male
  • Mitral Valve / diagnostic imaging*
  • Models, Animal
  • Reproducibility of Results
  • Swine
  • Ventricular Function, Left / physiology*
  • Ventricular Pressure / physiology*