Early hemodynamic performance of the Trifecta™ surgical bioprosthesis aortic valve in Indian patient population: 12 month outcomes of the EVEREST post-market study

J Cardiothorac Surg. 2018 Sep 25;13(1):96. doi: 10.1186/s13019-018-0783-9.

Abstract

Background: Indian patients undergoing surgical aortic valve replacement (SAVR) differ from western populations with respect to aortic annulus size and valve disease morphology. The purpose of this post-market, non-randomized observational study was to evaluate the early hemodynamic performance of the Trifecta™ bioprosthesis (Abbott, previously St. Jude Medical, Minneapolis, US) in an Indian patient population.

Methods: From January 2014 to September 2015, 100 patients (mean age 64.4 ± 7.1 years, 62% male) undergoing SAVR for valve disease (68% stenosis, 7% insufficiency, 25% mixed pathology) were enrolled across 10 centers in India. Patients implanted with a 19-27 mm Trifecta™ valve were eligible to participate and were prospectively followed for 12-months post-implantation. Echocardiographic hemodynamic performance was evaluated at pre-implant, pre-discharge and at 12-months by an independent core laboratory. Adverse events were adjudicated by the study sponsor. Functional status at 12-months was assessed according to NYHA classification. Continuous data was summarized using descriptive statistics (mean &standard deviation,) and categorical data was summarized using frequencies and percentages.

Result: Ninety patients (mean age 64.5, 62.2% male) completed the 12-month follow up. Significant improvements in hemodynamic valve performance were reported in 81 patients with available echocardiographic data at 12 months. Compared to baseline at 12-month follow up visit, mean effective orifice area increased from 0.75cm2 to 1.61cm2 (p < 0.0001), mean pressure gradient reduced to 10.42 mmHg from 51.47 mmHg (p < 0.0001), cardiac output increased from 4.46 l/min to 4.85 l/min (P 0.9254). Compared to baseline, functional status improved by ≥1 NYHA class in 75% of patients at 12 months (95% Clopper-Pearson (Exact) confidence limit [64.6%, 83.6%]). No instances of early mortality (< 30 days from index procedure) or structural valve dysfunction were reported.

Conclusion: In an Indian patient population, implantation of the Trifecta™ bioprosthesis is shown to be safe and associated with favorable early hemodynamic performance and improved functional status at 12 months.

Trial registration: The clinical study has been registered under Clinical Trial Registry-India ( http://www.ctri.nic.in ) and registration number is CTRI/2014/02/004434 registered on 25 February 2014 retrospectively registered.

Keywords: Aortic valve replacement; Stented bioprosthesis; Structural valve dysfunction; Valvular disease.

Publication types

  • Observational Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Insufficiency / physiopathology
  • Aortic Valve Insufficiency / surgery
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery
  • Bioprosthesis*
  • Echocardiography
  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis Implantation / methods*
  • Heart Valve Prosthesis*
  • Hemodynamics / physiology
  • Humans
  • Male
  • Middle Aged
  • Retrospective Studies
  • Time Factors