Influence of magnesium chloride on the dose-response of polyacrylamide-type gel dosimeters

Radiol Phys Technol. 2018 Dec;11(4):375-381. doi: 10.1007/s12194-018-0473-2. Epub 2018 Sep 4.

Abstract

We investigated the effect of magnesium chloride (MgCl2) on the nuclear magnetic resonance dose-response of polyacrylamide-type (PAGAT, NIPAM, and VIPET) gel dosimeters containing acrylamide, N-isopropylacrylamide, and N-vinylpyrrolidone as a monomer, respectively. The dose-transverse relaxation rates (1/T2 = R2) obtained from magnetic resonance imaging data revealed that a substantial increase in the dose-R2 response occurred as the concentration of MgCl2 in the gel dosimeters increased. The sensitivity of the PAGAT gel with 1.0 M MgCl2 was found to be approximately one order higher than that of the same gel without MgCl2. In addition, the water equivalences of the gels with MgCl2 were evaluated over a wide range of photon energies. The results indicated that MgCl2 acts as a powerful sensitizer to radiation-induced free-radical polymerization in polyacrylamide-type gel dosimeters, but does not interfere with the desirable properties of basic polyacrylamide-type gel dosimeters (i.e., the dose rate and dose integration).

Keywords: Magnesium chloride; NIPAM; PAGAT; Polymer gel dosimeter; Radiotherapy; VIPET.

MeSH terms

  • Acrylic Resins / chemistry*
  • Gels
  • Magnesium Chloride / chemistry*
  • Magnetic Resonance Imaging
  • Photons
  • Radiometry / instrumentation*