Micro-CT imaging as a method for comparing perfusion in graduated-height and single-height surgical staple lines

Med Devices (Auckl). 2018 Aug 15:11:267-273. doi: 10.2147/MDER.S171357. eCollection 2018.

Abstract

Background: Wound healing is a goal for advanced technology in the surgical space to benefit clinical outcomes. Surgical staplers are commonly used in a variety of open and minimally invasive abdominal and thoracic procedures. Assessment of wound healing traits, such as perfusion, has been challenging due to technical limitations. A novel technique that utilizes micro-computed tomography methodology to measure perfusion was designed to compare the micro-perfusion of staple lines between commercial stapler reloads that employ different staple height strategies.

Materials and methods: Following an Institutional Animal Care and Use Committee-approved protocol, rats were euthanized and immediately heparinized prior to a subtotal gastrectomy with either graduated-height or single-height staples. Rats were then perfused with barium, following which stomachs were removed and immediately fixed in formalin to prevent degradation. Stomachs were then imaged using micro-computed tomography and subsequent analysis was utilized to quantify fluid volume and patent vasculature proximity to staples within the staple line region for each group.

Results: Average perfusion volume was significantly higher with graduated-height staples (0.33% ± 0.18%) compared to single-height staples (0.16% ± 0.09%, P=0.011). Average vessel-to-staple line distance was not significant but trended lower with graduated-height staples (0.35±0.02 mm) compared to single-height staples (0.36±0.03 mm, P=0.18).

Discussion: Graduated-height staples had significantly higher perfusion volume than single-height staples, which likely has a downstream benefit on wound healing and clinical outcomes.

Conclusion: This study shows a higher perfusion volume around the staple lines using graduated-height staples as compared to single-height staples and this may contribute to better wound healing in patients.

Keywords: Tri-Staple; image processing; laparoscopic surgery; micro-computed tomography; microvasculature; wound healing.