Use of a steerable needle for CT-guided nerve plexus blockade

Abdom Radiol (NY). 2019 Jan;44(1):327-332. doi: 10.1007/s00261-018-1721-y.

Abstract

Purpose: Demonstrate the use of a new steerable needle for CT-guided neural plexus blockade to avoid traversing the kidneys.

Materials and methods: Retrospective evaluation of 23 consecutive CT-guided neural plexus blockade procedures in which a new steerable needle was used in the last 13 and compared with the prior 10 procedures in which a standard needle was used.

Results: Use of the steerable needle was beneficial to reach the target area without traversing a kidney or other organs in 6/13 (46%) procedures; in the other 7 procedures there was no benefit. A kidney was traversed in 0/13 procedures performed with the steerable needle. In contrast, a kidney was traversed in 4/10 (40%) procedures using a standard needle (P = 0.02). There was no significant difference in clinical benefit (P = 1.00) or complications (P = 0.56) between procedures using the steerable needle versus a standard needle. Three complications were observed (1 major and 2 minor) felt to be related to the injection and not the needle type.

Conclusions: The utility of a steerable 21-gauge needle during neural plexus blockades was found to allow for avoidance of the kidneys when compared to a standard (non-steerable) needle. Interventional radiologists may find this needle and its future iterations useful for neural blockades, as well as other procedures, when intervening structures need to be avoided.

Level of evidence: Level 3, Non-randomized controlled cohort.

Keywords: Nerve block; Pain management; Palliation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Equipment Design
  • Female
  • Humans
  • Male
  • Middle Aged
  • Needles*
  • Nerve Block / instrumentation*
  • Radiography, Interventional / methods*
  • Retrospective Studies
  • Tomography, X-Ray Computed / methods*
  • Young Adult