Meta-analysis of the incidence of lead dislodgement with conventional and leadless pacemaker systems

Pacing Clin Electrophysiol. 2018 Oct;41(10):1365-1371. doi: 10.1111/pace.13458. Epub 2018 Aug 27.

Abstract

Introduction: Leadless cardiac pacemaker (LCP) implantation using a transcatheter was recently developed to avoid pocket- and lead-related complications. Although a LCP has an active fixation mechanism using tines or a helix, LCP and lead dislodgement issues remain a major safety concern for patients. This article reviews the literature to determine the incidence of lead and LCP dislodgement.

Methods and results: A total of 18 studies which included 17,321 patients undergoing conventional single- or dual-chamber pacemaker implantation and three studies which included 2,116 patients undergoing LCP device implantation were reviewed. The incidence of lead dislodgement ranged from 1% to 2.69% in individual studies with a mean of 1.63%, weighted mean of 1.71%, and median of 1.60 %. There was a relatively higher lead dislodgement rate between atrial and ventricular electrodes (odds ratio [OR], 3.56; 95% confidence interval [CI], 1.9-6.70; P = 0.6; I2 = 0%), and between magnetic resonance imaging conditional and conventional leads (OR, 2.79; 95% CI, 1.30-5.99; P = 0.16; I2 = 46%). The use of active fixation leads (OR, 1.06; 95% CI, 0.66-1.70; P = 0.29; I2 = 20%) showed no significant difference in dislodgement risk compared to passive fixation leads. The incidence of LCP device dislodgement was 0%, 0.13%, and 1% in three leadless pacemaker studies.

Conclusions: The incidence rates of conventional pacemaker lead dislodgement vary in individual studies with an overall high incidence. Use of the currently available LCP systems appears to result in a lower rate of device dislodgement. This may reflect the effectiveness of this novel technology and the fixation design of LCP devices.

Keywords: cardiac pacemaker; incidence; lead dislodgement; leadless cardiac pacemaker; risk.

Publication types

  • Meta-Analysis
  • Review

MeSH terms

  • Equipment Design
  • Equipment Failure*
  • Humans
  • Incidence
  • Pacemaker, Artificial / adverse effects*
  • Risk Factors