Retrospective analysis of single-center early and midterm results of transapical catheter-based mitral paravalvular leak closure with a purpose-specific device

Postepy Kardiol Interwencyjnej. 2018;14(2):167-175. doi: 10.5114/aic.2018.76408. Epub 2018 Jun 19.

Abstract

Introduction: Due to the recent lack of definitions to establish the severity of paravalvular leak (PVL) and endpoints for its treatment, the effectiveness and safety of a new device for PVL closure have not been comprehensively analyzed.

Aim: To analyze a single center's experience of mitral PVL closure in a surgical transapical catheter-based fashion with a purpose-specific device.

Material and methods: This is a retrospective cohort study of patients following transapical catheter-based mitral PVL closure with a purpose-specific device. Data were analyzed at baseline, perioperatively, at discharge, at six months and annually after the procedure.

Results: Nineteen patients underwent surgical transapical catheter-based mitral PVL closure with the Occlutech PLD Occluder. Mean follow-up time was 20 ±7 (range: 9-33) months. The patients' mean age was 64 ±7 years, and 11 (58%) were male. Technical, device and individual patient success at follow-up was achieved in 18 (95%), 16 (84%) and 16 (84%) patients respectively. Median intensive therapy unit stay was one day (1-4) and mean hospital stay was 11 ±4 days. A reduction of paravalvular regurgitation to a mild or lesser degree was achieved in 18 (95%) patients. There were no strokes or myocardial infarctions at follow-up. There were no deaths at 30 days after the procedure. One (5%) patient expired due to progression of heart failure 12 months after surgery. None of the patients required immediate conversion to full sternotomy.

Conclusions: Surgical transapical catheter-based mitral PVL closure with the Occlutech PLD Occluder is a safe and clinically effective treatment.

Keywords: heart failure; mitral regurgitation; paravalvular leak; prosthetic heart valve; transcatheter closure.