Optimized Echo Decorrelation Imaging Feedback for Bulk Ultrasound Ablation Control

IEEE Trans Ultrason Ferroelectr Freq Control. 2018 Oct;65(10):1743-1755. doi: 10.1109/TUFFC.2018.2847599. Epub 2018 Jun 14.

Abstract

Feasibility of controlling bulk ultrasound (US) thermal ablation using echo decorrelation imaging was investigated in ex vivo bovine liver. The first of two ablation and control procedures used a sequence of constant-intensity sonication cycles, ceased when the minimum echo decorrelation within a control region of interest (ROI) exceeded a predetermined threshold. The second procedure used a variable-intensity sonication sequence, with spatially averaged decorrelation as the stopping criterion. US exposures and echo decorrelation imaging were performed by a linear image-ablate array. Based on preliminary experiments, control ROIs and thresholds for the minimum-decorrelation and average-decorrelation criteria were specified. Controlled trials for the minimum-decorrelation and average-decorrelation criteria were compared with uncontrolled trials employing 9 or 18 cycles of matching sonication sequences. Lesion dimensions, treatment times, ablation rates, and areas under receiver operating characteristic curves were statistically compared. Successfully controlled trials using both criteria required significantly shorter treatment times than corresponding 18-cycle treatments, with better ablation prediction performance than uncontrolled 9-cycle and 18-cycle treatments. Either control approach resulted in greater ablation rate than corresponding 9-cycle or 18-cycle uncontrolled approaches. A post hoc analysis studied the effect of exchanging control criteria between the two series of controlled experiments. For either group, the average time needed to exceed the alternative decorrelation threshold approximately matched the average duration of successfully controlled experimental trials. These results indicate that either approach, using minimum-decorrelation or average-decorrelation criteria, is feasible for control of bulk US ablation. In addition, use of a variable-intensity sonication sequence for bulk US thermal ablation can result in larger ablated regions compared to constant-intensity sonication sequences.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Algorithms
  • Animals
  • Cattle
  • Equipment Design
  • Feedback
  • High-Intensity Focused Ultrasound Ablation / methods*
  • Image Processing, Computer-Assisted / methods*
  • Liver / diagnostic imaging
  • Liver / surgery
  • Ultrasonography / methods*