Small-animal, whole-body imaging with metamaterial-inspired RF coil

NMR Biomed. 2018 Aug;31(8):e3952. doi: 10.1002/nbm.3952. Epub 2018 Jun 26.

Abstract

Particular applications in preclinical magnetic resonance imaging require the entire body of an animal to be imaged with sufficient quality. This is usually performed by combining regions scanned with small coils with high sensitivity or long scans using large coils with low sensitivity. Here, a metamaterial-inspired design employing a parallel array of wires operating on the principle of eigenmode hybridization was used to produce a small-animal imaging coil. The coil field distribution responsible for the coil field of view and sensitivity was simulated in an electromagnetic simulation package and the coil geometrical parameters were optimized for whole-body imaging. A prototype coil was then manufactured and assembled using brass telescopic tubes with copper plates as distributed capacitance. Its field distribution was measured experimentally using the B1+ mapping technique and was found to be in close correspondence with the simulated results. The coil field distribution was found to be suitable for large field of view small-animal imaging and the coil image quality was compared with a commercially available coil by whole-body scanning of living mice. Signal-to-noise measurements in living mice showed higher values than those of a commercially available coil with large receptive fields, and rivalled the performance of small receptive field and high-sensitivity coils. The coil was deemed to be suitable for some whole-body, small-animal preclinical applications.

Keywords: Acquisition Methods; Applications, Computational electromagnetics; Body; MR Engineering; Methods and Engineering; RF receive coils; RF transmit coils.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Computer Simulation
  • Kidney / diagnostic imaging
  • Magnetic Resonance Imaging / instrumentation*
  • Mice, Inbred BALB C
  • Radio Waves*
  • Signal-To-Noise Ratio
  • Whole Body Imaging*