Use of the iTClamp versus standard suturing techniques for securing chest tubes: A randomized controlled cadaver study

Turk J Emerg Med. 2018 Mar 9;18(1):15-19. doi: 10.1016/j.tjem.2018.01.006. eCollection 2018 Mar.

Abstract

Objectives: Tube thoracostomy (TT) is a common yet potentially life-saving trauma procedure. After successful placement however, securing a TT through suturing is a skillset that requires practice, risking that the TT may become dislodged during prehospital transport. The purpose of this study was to examine if the iTClamp was a simpler technique with equivalent effectiveness for securing TTs.

Materials and methods: In a cadaver model, a 1.5 inch incision was utilized along the upper border of the rib below the 5th intercostal space at the anterior axillary line. TTs (sizes 28Fr, 32Fr, 36Fr and 40Fr) were inserted and secured with both suturing and iTClamp techniques according to the preset randomization. TT were then functionally tested for positive and negative pressure as well as the force required to remove the TT (pull test-up to 5 lbs). Time to secure the TT was also recorded.

Results: When sutured is placed by a trained surgeon, the sutures and iTClamp were functionally equivalent for holding a positive and negative pressure. Mean pull force for both sutures and iTClamp exceeded the 5 lb threshold; there was no significant difference between the groups. Securing the TT with the iTClamp was significantly faster (p < 0.0001) with the iTClamp having a mean application time of 37.0 ± 22.8 s and using a suture had a man application time of 96.3 ± 29.0 s.

Conclusion: The iTClamp was effective in securing TTs. The main benefit to the iTClamp is that minimal skill is required to adequately secure a TT to ensure that it does not become dislodged during transport to a trauma center.

Keywords: Chest tube; Securing chest tubes; Tube thoracostomy.