In vivo performances of pure Zn and Zn-Fe alloy as biodegradable implants

J Mater Sci Mater Med. 2018 Jun 25;29(7):94. doi: 10.1007/s10856-018-6096-7.

Abstract

The disadvantage of current biodegradable metals such as Mg and Fe is the release of hydrogen gas in vivo that can cause gas embolism and the production of voluminous iron oxide that can cause inflammation, respectively. Such considerations have turned focus towards Zn as an alternative. This is based on the fact that Zn plays a crucial role in many physiological processes, as well as potentially being biocompatible and capable of with biodegradation. As such, the purpose of the present study was to evaluate the in vivo performance of pure Zinc and Zn-2%Fe implants. The use of iron as an alloying element was aimed at accelerating the corrosion rate of pure zinc by a micro-galvanic effect so as to maintain the post-implantation biodegradation characteristics of the implant. In vivo assessment was carried out using cylindrical disks implanted in the back midline of 16 male Wistar rats for up to 24 weeks. Post-implantation evaluation included monitoring the well-being of rats, weekly examination of hematological parameters: serum Zn levels, red and white blood cell counts and hemoglobin levels, X-ray radiography, histological analysis and corrosion rate assessment. The results obtained in terms of well-being, hematological tests and histological analysis of the rats indicate that the in vivo behavior of pure Zn and Zn-2%Fe implants was adequate and in line with the results obtained by the control group containing inert Ti-6Al-4V alloy implants. The corrosion rate of Zn-2%Fe alloy in in vivo conditions was relatively increased compared to pure Zn due to micro-galvanic corrosion.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Absorbable Implants* / adverse effects
  • Alloys / chemistry
  • Animals
  • Biocompatible Materials / adverse effects
  • Biocompatible Materials / chemistry
  • Corrosion
  • Iron* / chemistry
  • Male
  • Materials Testing
  • Rats
  • Rats, Wistar
  • Titanium / chemistry
  • Zinc* / blood
  • Zinc* / chemistry

Substances

  • Alloys
  • Biocompatible Materials
  • titanium alloy (TiAl6V4)
  • Titanium
  • Iron
  • Zinc