Ultra-low-dose computed tomography for attenuation correction of cadmium-zinc-telluride single photon emission computed tomography myocardial perfusion imaging

J Nucl Cardiol. 2020 Feb;27(1):228-237. doi: 10.1007/s12350-018-1303-y. Epub 2018 Jun 19.

Abstract

Background: The applicability of ultra-low-dose computed tomography (CT) for attenuation correction (AC) of single-photon-emission computed tomography myocardial perfusion imaging (SPECT-MPI) remains elusive.

Methods and results: One-hundred patients underwent one-day 99mTc-tetrofosmin stress-rest MPI and non-contrast enhanced cardiac CT with 120, 80, and 70 kilovolt peak (kVp) tube voltage and tube current of 200 milliamperes for creation of AC maps. Normalized percent myocardial uptake from SPECT-MPI using 80 kVp scans for AC showed excellent correlation vs AC from 120 kVp scans for stress [intraclass correlation (ICC) = 0.988, 95% CI = 0.986-0.989, P < .001] and rest (ICC = 0.985, 95% CI = 0.983-0.987, P < .001) with narrow Bland-Altman limits of agreement (BA-LA) (- 5.3% to 4.5% and - 5.4% to 4.4%, respectively) and minimal bias (- 0.4% and - 0.5%, respectively). Correlation of AC SPECT-MPI based on 70 vs 120 kVp scans was excellent for stress (ICC = 0.988, 95% CI = 0.986-0.989, P < .001) and rest (ICC = 0.986, 95% CI = 0.984-0.987, P < .001) with narrow BA-LA (- 5.3% to 4.4% and - 5.2% to 4.5%, respectively) and small bias (- 0.4% and - 0.3%, respectively). Mean effective radiation dose for the 120, 80 and 70 kVp scans were 0.58 ± 0.07, 0.19 ± 0.02, and 0.12 ± 0.01 mSv, respectively.

Conclusions: Attenuation maps for MPI obtained from ultra-low radiation dose CT scans are interchangeable with attenuation maps from standard-dose CT while offering a substantial reduction in radiation dose exposure.

Keywords: SPECT; attenuation correction; computed tomography; ultra-low dose.

MeSH terms

  • Aged
  • Cadmium
  • Female
  • Heart Diseases / diagnostic imaging*
  • Humans
  • Male
  • Middle Aged
  • Myocardial Perfusion Imaging*
  • Organophosphorus Compounds
  • Organotechnetium Compounds
  • Prospective Studies
  • Radiation Dosage
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tellurium
  • Tomography, Emission-Computed, Single-Photon*
  • Tomography, X-Ray Computed*
  • Zinc

Substances

  • CdZnTe
  • Organophosphorus Compounds
  • Organotechnetium Compounds
  • Cadmium
  • technetium tc-99m tetrofosmin
  • Zinc
  • Tellurium