Time to consider the contact force during photoplethysmography measurement during pediatric anesthesia: A prospective, nonrandomized interventional study

Paediatr Anaesth. 2018 Jul;28(7):660-667. doi: 10.1111/pan.13415. Epub 2018 Jun 19.

Abstract

Background: Respiratory variations in photoplethysmography amplitude enable volume status assessment. However, the contact force between the measurement site and sensor can affect photoplethysmography waveforms. We aimed to evaluate contact force effects on respiratory variations in photoplethysmography waveforms in children under general anesthesia.

Methods: Children aged 3-5 years were enrolled. After anesthetic induction, mechanical ventilation commenced at a tidal volume of 10 mL/kg. Photoplethysmographic signals were obtained in the supine position from the index finger using a force sensor-integrated clip-type photoplethysmography sensor that increased the contact force from 0-1.4 N for 20 respiratory cycles at each force. The AC amplitude (pulsatile component), DC amplitude (nonpulsatile component), AC/DC ratio, and respiratory variations in photoplethysmography amplitude were calculated.

Results: Data from 34 children were analyzed. Seven contact forces at 0.2-N increments were evaluated for each patient. The normalized AC amplitude increased maximally at a contact force of 0.4-0.6 N and decreased with increasing contact force. However, the normalized DC amplitude increased with a contact force exceeding 0.4 N. ΔPOP decreased slightly and increased from the point when the AC amplitude started to decrease as contact force increased. In a 0.2-1.2 N contact force range, significant changes in the normalized AC amplitude, normalized DC amplitude, AC/DC ratio, and respiratory variations in photoplethysmography amplitude were observed.

Conclusion: Respiratory variations in photoplethysmography amplitude changed according to variable contact forces; therefore, these measurements may not reflect respiration-induced stroke volume variations. Clinicians should consider contact force bias when interpreting morphological data from photoplethysmography signals.

Keywords: child; contacting force; oximetry; plethysmography.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Anesthesia, General
  • Biomechanical Phenomena
  • Child, Preschool
  • Female
  • Humans
  • Male
  • Photoplethysmography / instrumentation*
  • Photoplethysmography / methods*
  • Photoplethysmography / statistics & numerical data
  • Pressure
  • Prospective Studies
  • Respiration*
  • Respiration, Artificial