Quality assessment of uterine artery Doppler measurement in first-trimester combined screening for pre-eclampsia

Ultrasound Obstet Gynecol. 2019 Feb;53(2):245-250. doi: 10.1002/uog.19116. Epub 2019 Jan 11.

Abstract

Objective: To assess the quality of mean uterine artery (UtA) pulsatility index (PI) measurement in a first-trimester pre-eclampsia screening program.

Methods: Consecutive women with a singleton pregnancy attending first-trimester screening for fetal chromosomal abnormalities also had combined screening for pre-eclampsia based on the Fetal Medicine Foundation (FMF) algorithm, at a large practice in Sydney, Australia, from May 2014 to February 2017. Distributions of mean UtA-PI multiples of the median (MoM) on a logarithmic scale were plotted in relation to the normal median with 95% CI for each operator and for each month. Central tendency and dispersion and cumulative sum charts were produced. Mean UtA-PI MoM values between 0.95 and 1.05 were considered ideal and those between 0.90 and 1.10 were considered acceptable. The screen-positive rates for preterm pre-eclampsia in different groups of sonographers according to their mean log10 UtA-PI MoM were calculated and compared using the chi-square test.

Results: A total of 21 010 women attended for first-trimester ultrasound and had screening for pre-eclampsia. The overall median UtA-PI MoM was 1.042 (interquartile range (IQR), 0.85-1.26). Of 46 sonographers, 42 (91.3%) performed more than 50 examinations and, of those, 41 (97.6%) measured UtA-PI within the acceptable range. Sonographers measuring UtA-PI MoM on average below 0.95 and those measuring it above 1.05 had, respectively, lower and higher screen-positive rates when compared with those with measurements within the 0.95-1.05 UtA-PI MoM interval (7.2% and 13.2% vs 11.2%, respectively, P < 0.001).

Conclusion: UtA Doppler is measured well among trained operators when following an established protocol. While slight variations are expected, systematic error in this measurement impacts on the screen-positive rate. Therefore, a quality control process should be in place and retraining of staff may be required. Copyright © 2018 ISUOG. Published by John Wiley & Sons Ltd.

Keywords: first trimester; pre-eclampsia; quality assessment; screening; uterine artery Doppler.

MeSH terms

  • Adult
  • Diagnostic Errors / prevention & control
  • Female
  • Humans
  • Mass Screening / standards
  • Pre-Eclampsia / diagnostic imaging*
  • Pre-Eclampsia / prevention & control
  • Pregnancy
  • Pregnancy Trimester, First
  • Prospective Studies
  • Pulsatile Flow*
  • Ultrasonography, Doppler, Color / standards*
  • Ultrasonography, Prenatal / standards
  • Uterine Artery / diagnostic imaging*
  • Uterine Artery / physiopathology