Attenuated Total Reflection Fourier Transform Infrared (ATR FT-IR) Spectromicroscopy Using Synchrotron Radiation and Micromachined Silicon Wafers for Microfluidic Applications

Appl Spectrosc. 2018 Dec;72(12):1781-1789. doi: 10.1177/0003702818785640. Epub 2018 Aug 23.

Abstract

A custom-designed optical configuration compatible with the use of micromachined multigroove internal reflection elements (μ-groove IREs) for attenuated total reflectance Fourier transform infrared (ATR FT-IR) spectroscopy and imaging applications in microfluidic devices is described. The μ-groove IREs consist of several face-angled grooves etched into a single, monolithic silicon chip. The optical configuration permits individual grooves to be addressed by focusing synchrotron sourced IR light through a 150 µm pinhole aperture, restricting the beam spot size to a dimension smaller than that of the groove walls. The effective beam spot diameter at the ATR sampling plane is determined through deconvolution of the measured detector response and found to be 70 µm. The μ-groove IREs are highly compatible with standard photolithographic techniques as demonstrated by printing a 400 µm wide channel in an SU-8 film spin-coated on the IRE surface. Attenuated total reflection FT-IR mapping as a function of sample position across the channel illustrates the potential application of this approach for rapid prototyping of microfluidic devices.

Keywords: ATR FT-IR; IR spectromicroscopy; IR-SM; Micromachined internal reflection elements; attenuated total reflection Fourier transform infrared; macro ATR; microfluidics; photolithography; synchrotron infrared radiation.