Evaluation of adenosine triphosphate test for cleaning assessment of gastroscopes and the effect on workload in a busy endoscopy center

Am J Infect Control. 2018 Oct;46(10):1110-1114. doi: 10.1016/j.ajic.2018.04.215. Epub 2018 May 18.

Abstract

Objective: Using adenosine triphosphate (ATP) tests to assess manual cleaning of gastroscopes and to determine the associated workload in a busy endoscopy unit.

Methods: Patient-used gastroscopes were sampled before and after cleaning to assess ATP levels, bioburden, and protein. Samples were collected by flushing 20 mL of sterile water through the biopsy port to the distal end. Time spent for reprocessing and performing the ATP test was recorded.

Results: Twenty-four samples were collected from 10 gastroscopes. After manual cleaning, 14/24 (58.3%) samples had no microbial growth (mean, 21 colony-forming units/cm2), and in 22/24 (91.7%) samples the protein was undetectable (mean, 0.04 µg/cm2). ATP test was above the cutoff (200 relative light units [RLU]) in 17/24 (70.8%) samples (mean, 498 RLU). After the second cleaning, 11/17 (64.7%) gastroscopes still failed the ATP test (mean, 321.2 RLU). The mean time spent to perform manual cleaning and ATP tests was 16 and 8 minutes, respectively. Hence, each test increased the length of time for cleaning plus testing cleanliness by 50%.

Conclusion: Further studies regarding the optimal cutoff for ATP tests are needed. ATP tests for cleaning monitoring are easy to perform and provide immediate feedback to the team. However, the increased workload needs to be considered.

Keywords: Gastroscope; Rapid cleaning test; Reprocessing medical devices; Workload.